40 WATT MEDICAL POWER SUPPLIES
DESCRIPTION
The PM40AL series of compact, open PCB constructed, AC-DC
switching power supplies are capable of delivering 40 watts of
continuous output power. They operate at 85 to 264 VAC
input voltage without the need of voltage selection. They are
ideally suited for use in medical equipment, safety systems and
monitoring equipment, not for patient contact.
PM40AL SERIES
(MDD)
FEATURES
Low safety ground leakage current
100% burn-in
W i d e i n p u t r a n g e 8 5 t o 2 6 4 VA C
Input surge current protection
Overvoltage protection
Overcurrent protection
Open PCB construction
INPUT SPECIFICATIONS
Input voltage
Input frequency
Input current
Leakage current
85 to 264 VAC
47 to 63 Hz
1.20Armsfor 115VAC
0.70Armsfor 230VAC
150uA max. @ 230VAC, 50Hz
90 uA max. @ 115VAC, 60Hz
Safety Standard Approvals
UL 2601-1
CSA C22.2 No. 601.1
TÜV EN60601-1
OUTPUT SPECIFICATIONS
Output voltage/currentSee rating chart
Total output power
Ripple and noise
Overvoltage protection
Overcurrent protection
Temperature coefficient
Transient response
40 watts maximum
1peak to peak maximum
Provided on output #1 only; set at
112 - 132of its nominal output
voltage
All outputs protected to short
circuit conditions
All outputs ±0.04/maximum
Maximum excursion of 4or
better on all models, recovering to
1of final value within 500us
after a 25step load change
ENVIRONMENTAL SPECIFICATIONS
Operating temperature0to +70
Storage temperature-40to +85
Relative humidity
5to 95non-condensing
Derating
Derate from 100at +50linearly to
50% at +70
GENERAL SPECIFICATIONS
Switching frequency
36KHz ±5KHz
Efficiency
70minimum on single output models
with Vo15V, 65% minimum on the
Hold-up time
others
10 msec minimum at 110VAC
Line regulation
±0.5maximum at full load
Inrush current
15 amps @ 115VAC, or 30 amps @
230VAC, at 25cold start
Withstand voltage
4000VAC from input to output
1500VAC from input to ground
500VAC from output to ground
MTBF
600,000 hours minimum at full load at
25ambient, calculated per
MIL-HDBK-217F
EMC Performance (EN60601-1-2: 2001)
EN55011
Class B conducted, class B radiated
FCC
Class B conducted, class B radiated
VCCI
Class B conducted, class B radiated
EN61000-3-2
Harmonic distortion, class A and D
EN61000-3-3
Line flicker
EN61000-4-2
ESD, ± 8KV air and ± 6KV contact
EN61000-4-3
Radiated immunity, 3V/m for 80~2500MHz
EN61000-4-4
Fast transient/burst, ± 2KV
EN61000-4-5
Surge, ± 1KV diff., ± 2KV com.
EN61000-4-6
Conducted immunity, 3Vrms
EN61000-4-8
Magnetic field immunity, 3A/m
EN61000-4-11
Voltage dips, 30% reduction for 500ms,
60% reduction for 100ms and >95%
reduction for 10ms
Tumbler Technologies + TRUMPower
3350 Scott Blvd., Bldg. 13, Santa Clara, California 95054, USA
Phone: 408-988-6616 sales@trumpower.com www.TRUMPower.com
 