datasheetbank_Logo
Integrated circuits, Transistor, Semiconductors Search and Datasheet PDF Download Site

MPS-45-12 View Datasheet(PDF) - Mean Well Enterprises Co., Ltd.

Part Name
Description
View to exact match
MPS-45-12
MEANWELL
Mean Well Enterprises Co., Ltd. MEANWELL
MPS-45-12 Datasheet PDF : 2 Pages
1 2
45W Single Output Medical Type
MPS-45 series
Features :
Universal AC input / Full range
Low leakage current 0.3mA
Protections: Short circuit / Overload / Over voltage
Cooling by free air convection
100% full load burn-in test
Fixed switching frequency at 45KHz
3 years warranty
SPECIFICATION
MODEL
MPS-45-3.3 MPS-45-5 MPS-45-7.5 MPS-45-12 MPS-45-13.5 MPS-45-15 MPS-45-24 MPS-45-27 MPS-45-48
DC VOLTAGE
3.3V
5V
7.5V
12V
13.5V
15V
24V
27V
48V
RATED CURRENT
8A
8A
5.4A
3.7A
3.3A
3A
1.9A
1.7A
1A
CURRENT RANGE
0 ~ 10.7A 0 ~ 10.5A 0 ~ 7A
0 ~ 4.4A 0 ~ 3.9A 0 ~ 3.5A 0 ~ 2.2A 0 ~ 1.95A 0 ~ 1.1A
RATED POWER
26.4W
40W
40.5W
44.4W
44.55W
45W
45.6W
45.9W
48W
OUTPUT
OUTPUT POWER (max.)
52W(+3.3V:35W)with 18CFM min. Forced air convection
RIPPLE & NOISE (max.) Note.2 80mVp-p 100mVp-p 100mVp-p 100mVp-p 100mVp-p 100mVp-p 100mVp-p 100mVp-p 100mVp-p
VOLTAGE ADJ. RANGE
2.97 ~ 3.63V 4.5 ~ 5.5V 6.75 ~ 8.25V 10.8 ~ 13.2V 12.2 ~ 14.85V 13.5 ~ 16.5V 21.6 ~ 26.4V 24.3 ~ 29.7V 43.2 ~ 52.8V
VOLTAGE TOLERANCE Note.3 3.0%
3.0%
3.0%
2.0%
2.0%
2.0%
2.0%
2.0%
2.0%
LINE REGULATION
1.0%
1.0%
1.0%
1.0%
1.0%
1.0%
1.0%
1.0%
1.0%
LOAD REGULATION
3.0%
3.0%
3.0%
2.0%
2.0%
2.0%
2.0%
2.0%
2.0%
SETUP, RISE TIME
800ms, 30ms/230VAC 1200ms, 30ms/115VAC at full load
HOLD UP TIME (Typ.)
50ms/230VAC 16ms/115VAC at full load
VOLTAGE RANGE
FREQUENCY RANGE
90 ~ 264VAC
47 ~ 440Hz
127 ~ 370VDC
INPUT
EFFICIENCY(Typ.)
AC CURRENT (Typ.)
65%
72%
75%
76%
76%
77%
78%
78%
78%
1.2A/115VAC 0.7A/230VAC
INRUSH CURRENT (Typ.)
LEAKAGE CURRENT
COLD START 17A/115VAC
<0.3mA / 264VAC
35A/230VAC
OVERLOAD
PROTECTION
OVER VOLTAGE
WORKING TEMP.
WORKING HUMIDITY
ENVIRONMENT STORAGE TEMP., HUMIDITY
TEMP. COEFFICIENT
VIBRATION
SAFETY STANDARDS
SAFETY & WITHSTAND VOLTAGE
EMC
(Note 4)
ISOLATION RESISTANCE
EMC EMISSION
EMC IMMUNITY
MTBF
OTHERS DIMENSION
PACKING
53 ~ 75W (3.3V:36 ~ 55W) rated output power
Protection type : Hiccup mode, recovers automatically after fault condition is removed
3.8 ~ 4.46V 5.75 ~ 6.75V 8.63 ~ 10.1V 13.8 ~ 16.2V 15.5 ~ 18.2V 17.25 ~ 20.25V 27.6 ~ 32.4V 31 ~ 36.45V
Protection type : Hiccup mode, recovers automatically after fault condition is removed
-10 ~ +60 (Refer to "Derating Curve")
20 ~ 90% RH non-condensing
-20 ~ +85 , 10 ~ 95% RH
0.04%/ (0 ~ 50
10 ~ 500Hz, 2G 10min./1cycle, period for 60min. each along X, Y, Z axes
ANSI/AAMI ES60601-1, TUV EN60601-1, IEC60601-1 approved
I/P-O/P:4KVAC I/P-FG:1.5KVAC O/P-FG:0.5KVAC
I/P-O/P, I/P-FG:100M Ohms / 500VDC / 25 / 70% RH
Compliance to EN55011 (CISPR11) Class B, EN61000-3-2,-3
Compliance to EN61000-4-2,3,4,5,6,8,11, EN60601-1-2, medical level, criteria A
366.1Khrs min. MIL-HDBK-217F (25 )
127*76*28mm (L*W*H)
0.18Kg; 72pcs/15.1Kg/1.35CUFT
55.2 ~ 64.8V
NOTE
1. All parameters NOT specially mentioned are measured at 230VAC input, rated load and 25 of ambient temperature.
2. Ripple & noise are measured at 20MHz of bandwidth by using a 12" twisted pair-wire terminated with a 0.1uf & 47uf parallel capacitor.
3. Tolerance : includes set up tolerance, line regulation and load regulation.
4. The power supply is considered a component which will be installed into a final equipment. The final equipment must be re-confirmed that it still meets
EMC directives. For guidance on how to perform these EMC tests, please refer to “EMI testing of component power supplies.”
(as available on http://www.meanwell.com)
5. Mounting holes M1 and M2 should be grounded for EMI purposes.
6. Heat Sink HS1,HS2 can not be shorted.
File Name:MPS-45-SPEC 2011-12-15
 

Share Link: 